WP10
Sensor for the measurement of hydrogen peroxide – especially for high concentrations.
- Measured variable(s): Hydrogen peroxide
- Calibration:
- DIN 38409-15 “Determination of hydrogen peroxide”
- ISO/DIS 7157 “Determination of hydrogen peroxide content – Titrimetric method“
- Interferences:
- Must not contain Cl₂
- Must not contain PES
- Must not contain O₃
- Must not contain sulphides
- Must not contain phenols in aqueous solution >3 %
- pH range: 2-11
- Pressure range: 0‒1 bar, no pressure surges and/or fluctuations
- Temperature range: 0‒45 °C
- Integrated automatic temperature compensation
- Response time: T90 approx. 8 min
- Flow rate: approx. 30 l/h, low flow dependence
- Shaft length: standard 175 mm (mA version up to 220 mm)
- Connection: standard 4-pin plug; mA version 2-pole terminal, M12 socket or modbus RTU with M12 socket
- Material: PVC-U, stainless steel 1.4571
Product Description
Sensor for the measurement of hydrogen peroxide – especially for high concentrations.
Wasserstoffperoxid_WP10_171107
Download