Sensor for the measurement of peracetic acid C₂H₄O₃, surfactants and conductive acids are tolerated.
- Measured variable(s): Peracetic acid
 - Calibration:
- DIN 38409-15 “Determination of hydrogen peroxide”
 - ISO/DIS 7157 “Determination of hydrogen peroxide content – Titrimetric method”
 
 - Interferences:
- ClO₂ increases the reading
 - H₂O₂ has a very minor influence on the reading, lowers the PES signal
 - O₃ increases the reading significantly
 
 - pH range: 1-6
 - Pressure range: 0‒1 bar, no pressure surges and/or fluctuations
 - Temperature range: 0‒60 °C
 - Integrated automatic temperature compensation
 - Run-in period during first start-up: 30‒180 min
 - Response time: T90 approx. 3.5 min at 10 °C, approx. 45 s at 50 °C
 - Flow rate: approx. 30 l/h, low flow dependence
 - Shaft length: standard 175 mm (mA version up to 220 mm)
 - Connection: standard 4-pin plug; mA version 2-pole terminal, M12 socket or modbus RTU with M12 socket
 - Material: PEEK, stainless steel 1.4571
 