P10
Sensor for the measurement of peracetic acid – surfactants and conductive acids are tolerated.
- Measured variable(s): peracetic acid
- Calibration:
- DIN 38409-15 “Determination of hydrogen peroxide”
- ISO/DIS 7157 “Determination of hydrogen peroxide content – Titrimetric method”
- Interferences:
- ClO₂ is measured at a factor of 1
- H₂O₂ does not interfere
- O₃ is measured at a factor of 2500
- pH range: 1-6
- Pressure range: 0‒1 bar, no pressure surges and/or fluctuations
- Temperature range: 0‒45 °C
- Sensor with automatic temperature compensation
- Response time: T90 approx. 1.5‒5 min, depending on model and temperature
- Flow rate: approx. 45 l/h, low flow dependence
- Shaft length: standard 175 mm (mA version up to 220 mm)
- Connection: standard 4-pin plug; mA version 2-pole terminal, M12 socket or modbus RTU with M12 socket
- Material: PVC-U, stainless steel 1.4571
Product Description
Sensor for the measurement of peracetic acid – surfactants and conductive acids are tolerated.
Peressigsaeure_P10_EN_160811